Özet
High-performance thin-layer chromatography plays an important role in the determination of lipid fractions. However, it requires expensive equipment for quantification of lipids spots. In this experiment, flatbed scanner with gel analysis software was employed to determine lipid fractions of egg yolk, with testing some modifications: optical resolution (300 vs. 600 dpi), lane selection method (manual vs. automatic), and background subtraction options (none, lane edge, minimum profile, rubber band, rolling ball, or valley to valley). Data show that using flatbed scanner with gel analysis software is economically feasible and yields reliable and reproducible results. The best chromatogram was achieved at 600 dpi resolution with no background subtraction, regardless of lane selection. © 2013 Akadémiai Kiadó, Budapest.
Anahtar Kelimeler
Flatbed scanner | High quality image | TL 120 software | Visible light
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayımlanan tam makale
Dergi Adı JPC-JOURNAL OF PLANAR CHROMATOGRAPHY-MODERN TLC
Dergi ISSN 0933-4173
Dergi Tarandığı Indeksler SCI-Expanded
Makale Dili İngilizce
Basım Tarihi 06-2013
Cilt No 26
Sayı 3
Sayfalar 202 / 208
Doi Numarası 10.1556/JPC.26.2013.3.1