Objectives: Six sigma process analysis is widely used to evaluate the performance of tests in clinical laboratories. This study aimed to compare the sigma values determined by using the total allowable error (TEa) goals of The Clinical Laboratory Improvement Amendments (CLIA) and The European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) to evaluate the performance of some hemogram parameters. Methods: The study was conducted in Kastamonu Teaching and Research HospiJanuary-June 2022. Leukocyte, erythrocyte, hemoglobin, hematocrit, and platelet parameters studied on the Sysmex XN-1000 (Kobe, Japan) hematology autoanalyzer were included in the study. The 6-month mean bias and coefficient of variation (CV) were obtained retrospectively from proficiency test and internal quality control results, respectively. Sigma values were determined with the formula ‘(TEa-bias)/CV … |