Evaluation of Freestyle Optium Neo H and TaiDoc-TD4235 point-of-care meters for measuring blood β-hydroxybutyrate concentration in sheep
 
Yazarlar (2)
Doç. Dr. Kenan Çağrı TÜMER Kastamonu Üniversitesi, Türkiye
Mehmet A. Kılınç Bingöl Üniversitesi, Türkiye
Makale Türü Açık Erişim Özgün Makale (SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale)
Dergi Adı Veterinary Clinical Pathology (Q3)
Dergi ISSN 0275-6382 Wos Dergi Scopus Dergi
Dergi Tarandığı Indeksler SCI-Expanded
Makale Dili İngilizce Basım Tarihi 03-2023
Cilt / Sayı / Sayfa 52 / 2 / 288–294 DOI 10.1111/vcp.13223
Makale Linki http://dx.doi.org/10.1111/vcp.13223
Özet
Background Accurate and rapid measurement of blood β‐hydroxybutyrate (β‐OHB) concentrations is critical to identify hyperketonemia in sheep. Objective The objective of this study was to test the performance of Freestyle Optium Neo H and TaiDoc‐TD4235 point‐of‐care (POC) devices for measuring blood β‐OHB concentrations in sheep. Methods Venous blood samples from 105 sheep were evaluated with Freestyle Optium Neo H and TaiDoc‐TD4235 meters. A reference approach was the laboratory measurement of serum BHB concentrations using the Randox D‐3 Hydroxybutyrate reagent kit. Data were analyzed using Passing‐ Bablok regression and Bland–Altman analysis. The diagnostic accuracy was assessed using serum β‐OHB concentration at a cut‐off value of 0.8 mmol/L. Results Passing‐Bablok regression analysis revealed an intercept of 0.066 (CI95: 0.018–0.125) and slope of 1.242 (CI95 …
Anahtar Kelimeler
bias | ewe | point-of-care device | pregnancy toxemia | sensitivity | specificity
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
Web of Science 7
Scopus 6
Google Scholar 7
Evaluation of Freestyle Optium Neo H and TaiDoc-TD4235 point-of-care meters for measuring blood β-hydroxybutyrate concentration in sheep