| Background Accurate and rapid measurement of blood β‐hydroxybutyrate (β‐OHB) concentrations is critical to identify hyperketonemia in sheep. Objective The objective of this study was to test the performance of Freestyle Optium Neo H and TaiDoc‐TD4235 point‐of‐care (POC) devices for measuring blood β‐OHB concentrations in sheep. Methods Venous blood samples from 105 sheep were evaluated with Freestyle Optium Neo H and TaiDoc‐TD4235 meters. A reference approach was the laboratory measurement of serum BHB concentrations using the Randox D‐3 Hydroxybutyrate reagent kit. Data were analyzed using Passing‐ Bablok regression and Bland–Altman analysis. The diagnostic accuracy was assessed using serum β‐OHB concentration at a cut‐off value of 0.8 mmol/L. Results Passing‐Bablok regression analysis revealed an intercept of 0.066 (CI95: 0.018–0.125) and slope of 1.242 (CI95 … |