img
Optimization of Analytical Parameters for Determination of Iron Nickeland Cobalt in Plants with RP HPLC      
Yazarlar
Harun Çiftçi
Kırşehir Ahi Evran Üniversitesi, Türkiye
Ali Ölçücü
Türkiye
Ahmet Özkaya
Türkiye
Öğr. Gör. Tacettin ÇİFTÇİ Öğr. Gör. Tacettin ÇİFTÇİ
Kastamonu Üniversitesi, Türkiye
Özet
This study optimized the analytical parameters such as combination of mobile phase, pH of mobile phase, attempt effect of strange ions and current rate for simultaneous determination of iron, nickel and cobalt. 4-(2-Pyridiylazo)resorcinol (PAR) ligand is used for metal chelates. The metal-PAR chelates were separated on a Luna C18 siloxane bonded phase and eluted within 7.4 min of Co-PAR, 9.6 min of Fe-PAR and 12.4 min of Ni-PAR with methanol-water-acetonitrile-(l × 10'2 mol L"1) sodium acetate (55:40:2.5:2.5 v/v) (pH 5.3). Detection limits of process for Fe- PAR, Ni-PAR and Co-PAR under optimized conditions (500 nm) were found as 4.45,3.26 and 1.86 μg L-1 respectively. The analysis of Fe, Ni and Co ions in plant samples under optimized conditions were carried out by HPLC.
Anahtar Kelimeler
4-(2-Pyridylazo) resorcinol | Cobalt | HPLC | Iron | Nickel | Trace element | UV-Visible
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayımlanan tam makale
Dergi Adı ASIAN JOURNAL OF CHEMISTRY
Dergi ISSN 0970-7077
Dergi Tarandığı Indeksler SCI-Expanded
Makale Dili Türkçe
Basım Tarihi 04-2009
Cilt No 21
Sayı 4
Sayfalar 2643 / 2652
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
WoS 12
SCOPUS 13
Google Scholar 17

Paylaş