img
Prompt gamma neutron activation analysis of boron with a 241Am-Be neutron source   
Yazarlar
Prof. Dr. Şeref TURHAN
Kastamonu Üniversitesi
H Yücel
A Demirbas
Özet
A prompt gamma neutron activation analysis (PGAA) setup installed at ANRTC has been used to analyze boron. It consists of a 22.6% REGe detector and a 740 GBq 241Am-Be neutron source moderated with water and paraffin. At the sample irradiation position, the thermal neutron fluence rate measured was 2.36. 104 n. m-2·s -1 and the corresponding Cd-ratio was 22 for gold monitor. The absolute detection efficiency in the range of 120-1500 keV was determined using 152Eu standard solution. The sensitivity and detection limit for standard boric acid samples has been determined. The boron content in boric acid prepared from Turkish borate ores is measured to be 15.91±0.46% wt.
Anahtar Kelimeler
Makale Türü Özgün Makale
Makale Alt Türü ESCI dergilerinde yayımlanan tam makale
Dergi Adı JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY
Dergi ISSN 0236-5731
Makale Dili İngilizce
Basım Tarihi 1-2004
Cilt No 262
Sayı 3
Sayfalar 661 / 664
Doi Numarası 10.1007/s10967-004-0489-4
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
WoS 9
SCOPUS 13
Prompt gamma neutron activation analysis of boron with a 241Am-Be neutron source

Paylaş