Doç. Dr. Kenan Çağrı TÜMER
Kastamonu Üniversitesi, Türkiye |
|
Firat Üniversitesi, Türkiye |
|
Veterinarmedizinische Universitat Wien, Avusturya |
| Özet |
| Objective To evaluate the performance of a point‐of‐care hemoglobin/hematocrit measurement device (POCHD), validated for human medicine, for measuring hematocrit in calves. Design Prospective study conducted between July 2022 and April 2023. Setting University teaching hospital. Animals One hundred thirty‐three calves <1 month old, suffering from various diseases. Interventions The microhematocrit method (MHM) and a POCHD were used to measure hematocrit in venous blood samples. The MHM served as the reference method. The analytical performance of the POCHD was evaluated using Bland–Altman plots and calculating the intraassay coefficient of variation and observed total error (TEobs). Diagnostic performance was assessed by implementing receiver operating characteristic curve analysis and Cohen's kappa statistic. Measurements and Main Results Bland–Altman plots showed a … |
| Anahtar Kelimeler |
| anemia | calves | dehydration | method comparison | validation |
| Makale Türü | Özgün Makale |
| Makale Alt Türü | SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale |
| Dergi Adı | Journal of Veterinary Emergency and Critical Care |
| Dergi ISSN | 1479-3261 Wos Dergi Scopus Dergi |
| Dergi Tarandığı Indeksler | |
| Dergi Grubu | Q3 |
| Makale Dili | İngilizce |
| Basım Tarihi | 11-2025 |
| Sayı | 1 |
| Doi Numarası | 10.1111/vec.70048 |