| Özet |
| Anemia is a common clinical condition in cats, and accurate measurement of hematocrit (HCT) and hemoglobin (Hb) is essential for diagnosis and monitoring. Conventional laboratory methods are reliable but may not be practical for critically ill patients. Point-of-care (POC) devices offer rapid results using minimal blood volume, however studies evaluating their accuracy in cats remain limited. This prospective study evaluated the analytical and diagnostic performance of the Mission Ultra Hemoglobin/Hematocrit Testing System in 118 client-owned cats. HCT and Hb values obtained with Mission Ultra were compared with reference methods. The microcentrifuge method and the desktop autoanalyzer served as the reference methods for HCT and Hb measurements, respectively. Bland-Altman plots, Passing-Bablok regression, and Receiver Operating Characteristic (ROC) analysis were used to assess agreement and … |
| Anahtar Kelimeler |
| Anemia | Feline | Method comparison | Point-of-care test |
| Makale Türü | Özgün Makale |
| Makale Alt Türü | SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale |
| Dergi Adı | Research in Veterinary Science |
| Dergi ISSN | 0034-5288 Wos Dergi Scopus Dergi |
| Dergi Tarandığı Indeksler | SCI-Expanded |
| Dergi Grubu | Q2 |
| Makale Dili | İngilizce |
| Basım Tarihi | 12-2025 |
| Cilt No | 197 |
| Sayı | 1 |
| Doi Numarası | 10.1016/j.rvsc.2025.105943 |
| Makale Linki | https://doi.org/10.1016/j.rvsc.2025.105943 |