Is Single Measurement Enough to Get a Reliable Result with Optical Coherence Tomography
Yazarlar (4)
Arzu Taskiran Comez
Çanakkale Onsekiz Mart Üniversitesi, Türkiye
Ilker Eser
Acibadem Kadiköy Hospital, Türkiye
Prof. Dr. Coskun Bakar Çanakkale Onsekiz Mart Üniversitesi, Türkiye
Dr. Öğr. Üyesi Barış KÖMÜR Çanakkale Onsekiz Mart Üniversitesi, Türkiye
Makale Türü Açık Erişim Özgün Makale (SCOPUS dergilerinde yayınlanan tam makale)
Dergi Adı Türk Oftalmoloji Dergisi
Dergi ISSN 2147-2661
Dergi Tarandığı Indeksler Open Access Journals (DOAJ), Gale/Cengage Learning, Index Copernicus, EBSCO Database, Scopus, Turkish Medline-National Citation Index, Turk Medline, Cinahl, TUBITAK/ULAKBIM, Turkish Citation
Makale Dili Türkçe Basım Tarihi 01-2012
Kabul Tarihi Yayınlanma Tarihi 15-01-2012
Cilt / Sayı / Sayfa 42 / 1 / 11–15 DOI 10.4274/tjo.42.680770
Makale Linki http://cms.galenos.com.tr/FileIssue/6/119/article/11-15.pdf
UAK Araştırma Alanları
Göz Hastalıkları
Özet
ResultsDiscussionABSTRACTPurpose:To evaluate the repeatability and reliability of retinal nerve fiber layer (RNFL) thickness measurements using optical coherence tomography (OCT).Material and Method:Two hundred sixty-six eyes of 135 subjects (glaucoma, glaucoma suspects and healthy) were included in this study. Three sequential inferior, superior, nasal and temporal RNFL thickness measurements were performed using Spectral OCT [Opko/OTI, Inc., Miami, FL] by one operator at one session without pupillary dilatation. The differences between these three measurements of each quadrant in each eye were compared in microns and percentages. Repeated measures analysis of variance was performed for statistical analysis. Reliability is measured by intraclass correlation coefficient (ICC) for each quadrant.Results:ICCs of all quadrants ranged between 0.77 and 0.92, with the measurements of nasal quadrant being the least reproducible and the inferior being the most reproducible of all quadrants. RNFL measurement errors over 20% were seen in 9.63% of nasal quadrant, 5.3% of temporal quadrant, 0.6% of superior quadrant and only 0.3% of inferior quadrant measurements.Discussion:In order to get more repeatable and reliable results with OCT, sequential measurements more than one should be considered. We believe that special attention is required in the analysis of data of nasal and temporal quadrants.(Turk J Ophthalmol 2012; 42: 11-5)
Anahtar Kelimeler
Glaucoma | Optical coherence tomography | Reliability | Repeatability | Retinal nerve fiber layer
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
Scopus 2
Google Scholar 3
Is Single Measurement Enough to Get a Reliable Result with Optical Coherence Tomography

Paylaş