Dinamik kontur tonometri ve santral korneal kalınlık ilişkisi
Yazarlar (4)
İlker Eser İstanbul Bilgi Üniversitesi, Türkiye
Ezgı Yılmaz
Arzu Taşkıran Çömez
Çanakkale Onsekiz Mart Üniversitesi, Türkiye
Makale Türü Özgün Makale (SCOPUS dergilerinde yayınlanan tam makale)
Dergi Adı Glokom-Katarakt Dergisi
Dergi Tarandığı Indeksler EBSCO, Index Copernicus, Scopus Index, TUBITAK/ULAKBIM, Turkish Citation
Makale Dili Türkçe Basım Tarihi 04-2009
Cilt / Sayı / Sayfa 4 / 4 / 216–220 DOI
UAK Araştırma Alanları
Göz Hastalıkları
Özet
Purpose To investigate the relationship between intraocular pressure (IOP) measured by Pascal Dynamic Contour Tonometry (DCT) and central corneal thickness (CCT). Materials and Methods A hundred-sixteen eyes of 58 glaucoma suspected or healthy subjects (23 male, 35 female) with a mean age of 48.2±13.3 years were included in the study. After CCTs of all eyes measured by Ultrasonic pachymeter (NidekUS-4000, Japan), IOP was measured by DCT (Ziemer Group, Switzerland), then, by Goldmann Applanation Tonometer (GAT),(Optilasa, Spain). The relationship between measurements of these two tonometers and CCT was evaluated for statistical significance by Pearson correlation analysis. Results The mean value of CCT measurements was found to be 541.1±41.0 (455-635) μm. Mean IOPs measured by DCT and GAT were; 19.6±3.3 (12.6-29.4) mmHg and 18.5±3.1 (12-28) mmHg, respectively …